| Name | HddtLungFunctionReferenceValue |
| Title | Observation – Lung Function Reference Value |
| Description | Profile for capturing the refence values as a FHIR Observation resource when evaluating lung function testings.
This profile defines the exchange of a single reference value for the Mandatory Interoperable Value (MIV) \"Lung Function Testing\" which is technically defined
by the ValueSet _hddt-miv-lung-function-testing_. This MIV is e.g. implemented by peak flow meter that can connect to
a Personal Health Gateway (e.g. a mobile app for tracking lung function values) through wireless or wired communication.
**Obligations and Conventions:**
Each Lung Function Testing MAY either hold a reference to a _Sensor Type And Calibration Status_ [DeviceMetric](https://hl7.org/fhir/R4/devicemetric.html) resource or to a
_Personal Health Device_ [Device](https://hl7.org/fhir/R4/device.html) resource (eXclusive OR). Typically the reference will be
to a [Device](https://hl7.org/fhir/R4/device.html) resource, but the option to reference a [DeviceMetric](https://hl7.org/fhir/R4/devicemetric.html)
resource is provided for compatibility with the overarching HDDT specification.
**Constraints applied:**
- `status` is restricted to _final_
- `code` is constrained to a subset of the _MIV Lung Function Reference Values_ ValueSet, defined by the _HddtLungFunctionReferenceValues_ ValueSet.
- `effective[x]` is restricted to `effectivePeriod` and constrained as mandatory.
- `value[x]` is restricted to `valueQuantity`. The elements `valueQuantity.value`, `valueQuantity.system`, and `valueQuantity.code` are constrained in a way that a value MUST be provided and that UCUM MUST be used for encoding the unit of measurement. `Observation.valueQuantity` MAY only be omitted in case of an error that accured with the measurement. In this case, `Observation.dataAbsentReason` MUST be provided.
- `method` is considered mandatory in order to provide information about the method used to determine the reference value. It can be either a code from the _HddtLungFunctionReferenceValueMethod_ ValueSet or a text description. |
| Copyright | Copyright (c) 2026 gematik GmbH |
| Type | Observation |
| Kind | resource |
Resources that use this resource
Resources that this resource uses
Narrative
No narrative content found in resource
Source1
{
"resourceType": "StructureDefinition",
"id": "hddt-lung-reference-value",
"url": "https://gematik.de/fhir/hddt/StructureDefinition/hddt-lung-reference-value",
"version": "1.0.0-rc",
"name": "HddtLungFunctionReferenceValue",
"title": "Observation – Lung Function Reference Value",
"status": "active",
"date": "2026-01-16",
"publisher": "gematik GmbH",
"description": "Profile for capturing the refence values as a FHIR Observation resource when evaluating lung function testings.\n\nThis profile defines the exchange of a single reference value for the Mandatory Interoperable Value (MIV) \\\"Lung Function Testing\\\" which is technically defined \nby the ValueSet _hddt-miv-lung-function-testing_. This MIV is e.g. implemented by peak flow meter that can connect to \na Personal Health Gateway (e.g. a mobile app for tracking lung function values) through wireless or wired communication.\n\n**Obligations and Conventions:**\n\nEach Lung Function Testing MAY either hold a reference to a _Sensor Type And Calibration Status_ [DeviceMetric](https://hl7.org/fhir/R4/devicemetric.html) resource or to a \n_Personal Health Device_ [Device](https://hl7.org/fhir/R4/device.html) resource (eXclusive OR). Typically the reference will be \nto a [Device](https://hl7.org/fhir/R4/device.html) resource, but the option to reference a [DeviceMetric](https://hl7.org/fhir/R4/devicemetric.html) \nresource is provided for compatibility with the overarching HDDT specification.\n\n**Constraints applied:** \n- `status` is restricted to _final_\n- `code` is constrained to a subset of the _MIV Lung Function Reference Values_ ValueSet, defined by the _HddtLungFunctionReferenceValues_ ValueSet.\n- `effective[x]` is restricted to `effectivePeriod` and constrained as mandatory.\n- `value[x]` is restricted to `valueQuantity`. The elements `valueQuantity.value`, `valueQuantity.system`, and `valueQuantity.code` are constrained in a way that a value MUST be provided and that UCUM MUST be used for encoding the unit of measurement. `Observation.valueQuantity` MAY only be omitted in case of an error that accured with the measurement. In this case, `Observation.dataAbsentReason` MUST be provided.\n- `method` is considered mandatory in order to provide information about the method used to determine the reference value. It can be either a code from the _HddtLungFunctionReferenceValueMethod_ ValueSet or a text description.",
"copyright": "Copyright (c) 2026 gematik GmbH",
"fhirVersion": "4.0.1",
"kind": "resource",
"abstract": false,
"type": "Observation",
"baseDefinition": "http://hl7.org/fhir/StructureDefinition/Observation",
"derivation": "constraint",
"differential": {
"element": [
{
"id": "Observation.status",
"path": "Observation.status",
"short": "Measurement status",
"definition": "The status of the measurements is fixed to 'final'. Only verified and complete measurements with a valid value are represented.",
"fixedCode": "final"
},
{
"id": "Observation.code",
"path": "Observation.code",
"short": "Reference measurement type for lung function",
"binding": {
"strength": "required",
"valueSet": "https://gematik.de/fhir/hddt/ValueSet/hddt-lung-function-reference-values",
"description": "Specifies for which measurement type the reference value is, using codes from the ValueSet for lung function testings. Constrained via invariant to either PEF or FEV1."
}
},
{
"id": "Observation.effective[x]",
"path": "Observation.effective[x]",
"short": "Time period for which reference value is valid",
"definition": "The time period for which the lung function reference value is valid. If the reference value is still active, then only the start of the period is provided.",
"type": [
{
"code": "Period"
}
]
},
{
"id": "Observation.value[x]",
"path": "Observation.value[x]",
"short": "Reference lung function value",
"definition": "The quantitative reference value, measured either in liters (L) for FEV1 or liters per minute (L/min) for PEF, represented as a UCUM Quantity.",
"type": [
{
"code": "Quantity"
}
],
"mustSupport": true
},
{
"id": "Observation.value[x].value",
"path": "Observation.value[x].value",
"min": 1
},
{
"id": "Observation.value[x].system",
"path": "Observation.value[x].system",
"min": 1,
"fixedUri": "http://unitsofmeasure.org"
},
{
"id": "Observation.value[x].code",
"path": "Observation.value[x].code",
"definition": "The UCUM code representing the unit of the lung function testing. The UCUM code MUST be compliant with the example unit that is linked with the LOINC code given as `Observation.code`.",
"min": 1,
"binding": {
"strength": "required",
"valueSet": "http://hl7.org/fhir/ValueSet/ucum-units",
"description": "Defines the unit of measurement using codes from the UCUM units ValueSet. The UCUM code MUST be compliant with the unit that is linked with the LOINC code given as `Observation.code`."
}
},
{
"id": "Observation.method",
"path": "Observation.method",
"short": "Method for determining the reference value",
"definition": "The method used to determine the lung function reference value. Preferred is the usage of a code from the Lung Function Reference Value Method ValueSet, but as an alternative a text description of the used formula for calculating the reference value can be provided.",
"mustSupport": true
},
{
"id": "Observation.method.coding",
"path": "Observation.method.coding",
"max": "1",
"binding": {
"strength": "required",
"valueSet": "https://gematik.de/fhir/hddt/ValueSet/hddt-lung-function-reference-value-method",
"description": "Specifies the method used to determine the lung function reference value using codes from the Lung Function Reference Value Method ValueSet."
}
},
{
"id": "Observation.method.text",
"path": "Observation.method.text",
"definition": "In case no code is provided, specify whether the value is a personal best or predicted or calculated based on some formula."
},
{
"id": "Observation.device",
"path": "Observation.device",
"short": "Reference to personal health device",
"definition": "References a Device resource that describes the personal health device.",
"type": [
{
"code": "Reference",
"targetProfile": [
"https://gematik.de/fhir/hddt/StructureDefinition/hddt-personal-health-device",
"https://gematik.de/fhir/hddt/StructureDefinition/hddt-sensor-type-and-calibration-status"
]
}
]
}
]
}
}